Ammonium Level Monitor

Camlab

The 230V FAM Ammonia PVC, an electrode based system from Swan, will facilitate continuous Ammonium level measurement in potable or effluent waters.

A transmitter and measuring cell are included on a PVC back-panel.

Optional Features

  • RS485 interface - a choice of PROFIBUS DP, SWANBUS, MODBUS ASCII/RTU protocol.
  • Potassium sensor compensates automatically for ion interference (potassium).

Ammonium Level MonitorTechnical Details

  • Transmitter housing IP66
  • Large backlit LCD screen for indicating operating status and measurement values
  • 2 signal outputs electrically isolated from sensor input, for temperature and process value 0 - 20mA
  • 2 inputs for potential-free switches with customisable functions
  • 2 potential free contacts, that can be programmed as PID controllers or limit switches
  • Sample flow 20 to 40l/h
  • Sample pressure range 0.15 to 2bar
  • Mounting panel dimensions 280 x 850mm
  • High voltage protection for signal outputs and RS485

Measurement Range Details

  • Automatic temperature compensation
  • Measuring range 0.1 to 1000ppm
  • Resolution 0.01ppm in the 0.00 to 9.99 range, 0.1ppm 10.0 to 99.9 and 1ppm 100 to 1000
  • Accuracy 10% of measured value
  • Temperature measurement -10 to +50°C, resolution 0.1°C.

£3,130.00 ex. VAT